Delta NANOTRACE DF-745 Instruction Manual Page 24

  • Download
  • Add to my manuals
  • Print
  • Page
    / 98
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 23
20 DF-745 Installation, Start Up and Shut Down
The moisture cell outlet will be connected to the supplied aspirator or, if equipped, to the
vacuum pump.
The sample inlet to the moisture analyzer, as well as the zero gas inlet, are heated to 60 C.
This is done to mitigate any effects of adsorption-desorption of trace moisture on the walls
of the tubing.
4.2.2 Gas Pressure and Flow Settings
Attention to the setting of gas pressure and flow is critical to proper operation of the
analyzer. If all steps are followed carefully at the time of start up, subsequent changes to
flow or background gas will be made easier. The following procedure assumes all electrical
and plumbing connections have been made according to instructions in this manual. In
addition, this procedure assumes a Nitrogen gas background unless otherwise noted.
1. Power up the unit. See page 19. If the unit is equipped with a Hydrogen Safety
System, no flow will enter the system until the unit is on power.
2. Ensure that the needle valve on the aspirator is fully open.
3. For a basic analyzer with no cell isolation valves installed,
a. simply provide sufficient pressure (5-30 psi), resulting in a flow of 1 slpm (2
scfh) as indicated on the flow meter located behind the front door. NOTE:
The flow rate will directly impact the overall system speed of response.
b. Turn on the vacuum source (aspirator) by applying gas to the high pressure
inlet. See Figure 3. (Except when measuring in Helium background which is
not run under vacuum but at ambient) NOTE: Once the vacuum source is
connected and running, proper flow is indicated by a full scale reading on the
flowmeter.
c. Go to step 5.
4. For analyzers equipped with optional cell isolation valves:
a. Completely open (turn fully clockwise) the inlet regulator on the rear of the
analyzer. For operation in Nitrogen, Argon and Oxygen the regulator is
closed (turned counter clockwise) to the approximate middle, or 50% of its
range. For operation in Helium and Hydrogen it should be closed to 80-90%
of its range resulting in a flow of 1 slpm (2 scfh) as indicated on the flow
meter located behind the front door. NOTE: The flow rate will directly
impact the overall system speed of response.
b. Open the flowmeter bypass valve on the rear of the analyzer and flow in the
bypass loop will be indicated on the flowmeter. Adjust the flowmeter bypass
valve to 0.5 slpm. See Figure 5.
c. Purge for 15 minutes before opening gas valves to the moisture cell.
d. The state of the gas control valves is indicated on the main display. The
default state of a factory-shipped instrument is isolation. This can be
adjusted in the Power Up Default section. If the moisture cell is isolated,
establish process flow via the Main menu. See page 38.
Page view 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 97 98

Comments to this Manuals

No comments