Delta NANOTRACE DF-745 Instruction Manual Page 45

  • Download
  • Add to my manuals
  • Print
  • Page
    / 98
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 44
User Interface DF-745 41
back and change the value or move to the Done button, followed by hitting the Enter key
to leave the screen. Using the ESC key at any time will exit the screen making no changes
and return to the main display.
7.4.3.1.2 Do A Manual Zero
Figure 29: Manual Zero Screen
The manual zero command enables the user to zero the moisture cell in an interactive
manner.
For this purpose, it is necessary to ensure that moisture free gas (sub ppb) is entering the
sensor through the process inlet. When switching to a gas that is moisture free, it is
important to wait a period necessary to allow the reading to re-stabilize.
It is extremely important to note that moisture is inherently slow to move and as a result the
process of doing a zero action can take many hours before the reading is at baseline. A
premature zero, although acceptable, will ultimately result in a negative reading as the
plumbing and analyzer continue to dry out requiring another zero action. See the section on
Active Zero on page 57 for additional information.
After selecting Do a Manual Zero, a screen will appear which displays a trace of the
recent moisture reading. See Figure 29. Observe the trace until the reading is stable and
then press Enter. This action will accept the present value as the new zero setting for the
moisture cell and the Zero Reference field will be updated. After this action, the user will
be brought back to the Check/Adjust Zero screen.
Pressing ESC during the calibration process will abort the action and return to the main
display. Whether the calibration is complete or aborted, pressing Done from the previous
screen will allow the user to return to the main display.
Once complete the gas source must be returned to the original state with process gas going
directly through the sensor.
A relay is available on the Analog Output Setup Screen (see page 55) to signal that a zero
calibration is taking place, and the analog output signal can also be frozen or allowed to
update during the calibration process.