Delta NANOTRACE DF-745 Instruction Manual Page 67

  • Download
  • Add to my manuals
  • Print
  • Page
    / 98
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 66
User Interface DF-745 63
7.4.12 Power Up Default Optional
Figure 71: Power Up Default Menu
This menu item is only available on analyzers equipped with the isolation valve option.
The Power Up Default SubRoutine allows the user to determine the various default states
during analyzer power up. The power up states are useful because they determine whether,
for instance, the sensor is protected from ambient air contamination or whether it is
configured the best way for rapid station to station transfer and measurement.
Use the NEXT key to move from field to field. Use the up and down arrows ( and )
to move between the selections within the field. When done use the NEXT key to move to
the Accept field and hit the Enter key. Pressing the ESC at any time will make no changes
and will return to the main data display.
Figure 72: Power Up Default Screen
NOTE: If the analyzer auto-reboots due to a system error 20006, the power-up defaults (if
applicable) are ignored and the analyzer returns to the mode of operation found immediately
before the error. This reboot action is identified in the system error log as an error 222. See
page 47 for additional information on system error codes.