Delta 4000 Instruction Manual Page 16

  • Download
  • Add to my manuals
  • Print
  • Page
    / 108
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 15
M
81331 Rev 01 Nov 2010
8
Power Factor: 0 to 100%, 0.001% maximum resolution
Dissipation Factor: 0 to 100, 0.001% maximum resolution
Watt Loss: 0 to 2 kW, actual power, 0 to 100 kW when
corrected to 10 kV equivalent. 0.1mW
maximum resolution.
The measurement can be corrected to either
2.5 kV or 10 kV equivalents.
Temperature
Correction:
Intelligent temperature correction (ITC)
from 5°C to 50°C test temperature to 20°C
reference
Standard tables for temperature corrections
Accuracy:
Voltage: ±(1% of reading + 1 digit)
Current: ±(1% of reading + 1 digit)
Capacitance: ±(0.5% of reading + 1 pF)
Inductance: ±(0.5% of reading + 1 mH)
Power Factor: ±(0.5% of reading + 0.02%)
Dissipation Factor: ±(0.5% of reading + 0.02%)
Watt Loss: ±(1% of reading + 1mW)
Noise Immunity:
Electrostatic: 15mA induced noise into any test lead with
no loss of measurement accuracy at
maximum interference to specimen current
of 20:1
Electromagnetic: 500 μT, at 50 Hz in any direction