Delta MMC120 User Manual Page 111

  • Download
  • Add to my manuals
  • Print
  • Page
    / 150
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 110
111
Click here for MDT SCALE Calculation examples
NOTE: When the SCALE parameter is changed, the EXTEND LIMIT, RETRACT
LIMIT,andOFFSET must also be changed.
NOTE:You should never set the SCALE below 16383. If it does go below that value,
then use the Prescale Divisor bits in the CONFIGURATION WORD.Asyou
can see from the ACTUAL POSITION calculation above, this will affect the
scaling, but will reduce round-off errors.
SCALE Calculation Examples
Example 1
For a system using a magnetostrictive transducer with a calibration number of
9.1392 µs per inch, a position unit of 0.001 inch, and 1 recirculation, the SCALE
will be:
scale = 1000(Position Units Per Inch) x 32768= 29878.62
9.1392(µs/in) x 120MHz x 1 (Recirculations)
SCALE = 29879
Example 2
For a system using a magnetostrictive transducer with a calibration number of
9.0110 µs per inch, a position unit of 0.1 millimeter, and 1 recirculation, the
SCALE will be:
scale = 254(Position Units Per Inch) x 32768 = 7697.14
9.0110(µs/in) x 120MHz x 1 (Recirculations)
Since the scale is less than 16383.5, use the Prescale Divisor bits in the
CONFIGURATION WORD to divide the transducer counts by 4. Then multiply
the scale by 4 to obtain:
SCALE = 7697.14x4=30789
Example 3
For a system using a magnetostrictive transducer with a calibration number of
8.9772 µs per inch, a position unit of 0.001 inch, and 8 recirculations, the SCALE
will be:
scale = 1000(Position Units Per Inch) x 32768 = 3802.22
8.9772(µs/in) x 120MHz x 8 (Recirculations)
Page view 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 149 150

Comments to this Manuals

No comments