User Manual | DELTAGuard | Trouble-Shooting
Error diagnostics fume cupboard sash control (ATTENTION! SHALL ONLY BE CONDUCTED BY SERVICE STAFF)
Error possible cause Remedy
Keys UP, DOWN do not
function
n 230V AC Voltage supply not
connected or faulty
n Defect connection cable between
key and electronics
n Defect motor cable
n Check voltage supply, check
electronics fuses, if needed,
replace them
n Check connection cable, if needed it
should be replaced
n Check motor cable, if needed it should
be replaced
Drive does not run
n 230V AC Voltage supply not
connected or faulty
n Defect connection cable between
key and electronics
n Defect motor cable
n Check voltage supply, check
electronics fuses, if needed,
replace them
n Check connection cable, if needed it
should be replaced
n Check motor cable, if needed it should
be replaced
Drive stops in an uncon-
trolled way when moving
upward or downward and
no obstacle is interrupting
the light barrier
n In case of a front slide distance of
more than 1 m a standard motor
unit is installed
n Polluted sender/receiver optics
n Damaged light barrier cable
n Light ray from sender to receiver
not adjusted
n Defect passive infrared sensor
(PIR)
n In case of a front slide distance of
more than 1 m the special version of
the motor drive unit shall be installed
n Clean sender/receiver optics
n Check light barrier cable, if needed it
should be replaced
n Adjust light ray
n Check sensor, if needed, adjust or
replace it
Automatic locking process
is not functioning
n Defect or incorrectly adjusted
passive infrared sensor (PIR)
n Defect relay output of the passive
infrared sensor (PIR)
n Polluted or defect sender/receiver
optics
n Check sensor, if needed it should
be replaced
n Check relay output, if needed it should
be repaired
n Check sender/receiver optics, if
needed it should be cleaned or repaired
6.2 Error diagnostics fume cupboard sash control
Attention! The error diagnostic of the fume cupboard sash control shall only be conducted by service staff.