DELTA LTL2000 Retrometer 15
Functions and Menu system
Rl Calibration
(Activate with the <Calib.> key)
LCD Display Explanation.
Rl Normal Cal.
Enter To Start
Performs instrument calibration.
<ENTER> opens Rl Calibration sub menu.
Mount Normal
Enter When Ready
The user is requested to place his Rl-Normal in the LTL2000
instrument.
Press <ENTER> when it is mounted correctly.
Calibrating
Please wait
Measures Rl on the mounted Rl-Normal.
Rl Normal 146
Enter=OKêé=edit
The display shows the measured Rl value.
Use <ENTER> or <ê><é> keys
Rl Normal 146
Enter When OK
If necessary use the <ê><é> keys to change the Rl reading down or
up, continue until the shown Rl normal value is identical to the value
printed on the calibration unit and press <ENTER>.
Calib Done
Remove Rl Normal
The display shows that a new Rl Normal value has been calculated.
The user is requested to remove the Rl normal.
Zero measurement
(Activate with the <2.nd.>+<ZERO> key)
LCD Display Explanation.
Zero Reading
Enter to Start
Ready for measuring the instrument ZERO signal. Make sure that the
light trap is mounted and empty, close the hatch and then press the
<ENTER> key.
Zero Reading
Please Wait
Measures the zero signal
Zero OK
Zero: 0.11%
Zero measurement done no errors detected. The measured signal level
is shown.
High Zero Warning
Zero: 7.61%
Warning condition detected, the signal is to high. Check that the light
trap is empty and clean. The following measurements will have the W
status set.
Zero OK
Ready
Instrument Zero OK, now ready for calibration.