DELTA LTL2000 Retrometer 19
Status
(Select with the <Menu> key)
LCD Display Explanation.
Status Display
Use éê to View
The results and status made with the Rl, Test, Calib. and the Zero
functions can be displayed in the LCD one at a time by use of the
<é><ê> keys.
The following messages are possible.
High Signal Warning
Signal Overflow
Signal = 12.4%
The measured signal strength in % of full scale. If the measured signal
is over a predefined limit then a warning will be displayed for one
second.
The measured signal level is shown
High Stray Light
Stray L = 0%
Stray L = 4,51%
No Stray L. Data
The measured stray light, representing the amount of light getting into
the LTL2000 from outside. A high Stray Light signal can indicate
problems with direct sunlight from a low sun on a very rough or
uneven surface. Special precautions must be taken to hinder the light
from shining directly on the measurement area. A warning is given in
the display if the signal gets critical.
Stray Light levels under a predefined limit are displayed as 0%
No Stay Light data after a ZERO measurement.
Low VBat Warning
VBat idle =10.6V
The battery voltage measured just before the lamp is lit. The reading
can be used to check the charge condition of the built-in 12 V Lead
Acid accumulator. A warning is given in the display if the voltage
drops below its predefined minimum limit.
Low VBat Warning
VBat load =11.2V
The battery voltage measured while the lamp is lit. The reading can be
used to check the charge condition of the built-in 12 V Lead Acid
accumulator. A warning is given in the display if the voltage under load
drops to a critical low value. If the battery voltage gets very low, the
LTL2000 will shut off during the measurement because the battery
voltage drops below its predefined minimum level when the lamp is lit.
This condition will be reported in the display on the next power Up.
Measurements will not be possible until the battery is recharged or an
external power source is applied.
Rl=146 mcd/m
2
/lx The latest Rl value measured.
Rl Log Full
or
Free Log: 96.1%
A warning is given if the Rl Log has run full. This means that it contains
approx. 1000 Rl readings, and that no new measurements can go into
the log. The log will have to be emptied and cleared before this
warning goes away. Also the measurement number in the printout will
not change until the log is cleared.