LTL2000 Command set
The LTL2000 is equipped with a serial communication port primarily for log data acquisition, calibration and test, however all normal LTL2000 functions can be
controlled from this interface.
The following commands have been defined
LTL2000 command set
Command Parameter Response Meaning
DA none
1997 3 27
1997 Mar 21 09:45:47
1997 Mar 27 16:39:45
Real time clock Date and time
New Date (time unchanged)
DR none *** Rl Measurement ***
1994-Mar-27 10:03:56
Rl: 145 (mcd/m
2
)/lx
Status Code: 0 00000000
Query Latest Rl Measurement
Date and time
Result
Status
DS none *** Zero Measurement ***
Lamp On : 12
Lamp Off : 10
System Leak : 3
*** Calibration ***
Lamp On : 163
Lamp Off : 10
On-(Off+Sys_Leak) : 151
Rl_Factor : 1.937
*** Rl, RT and Internal Data ***
Lamp Off : 149
Stray Light : 0
Rl_Result : 144.6
Quick overview of the latest measurements
Latest zero measurement
Signal with lamp on
Signal with lamp off
System leak signal
Latest Calibration
Signal with lamp on
Signal with lamp off
Calculated signal
Calculated Rl_Factor
Latest Rl, RT measurement
Signal with lamp off
Stray light signal
Rl result
DT none *** Rl Test Measurement ***
1997-Mar-27 10:02:30
Rl =144 (mcd/m
2
)/lx
Signal: 7.3%
Stray Light 0.0%
Latest test result
Date and Time
Measured Rl result
Measured signal
Measured stray light signal