DELTA LTL2000 Retrometer 24
LTL2000 command set
Command Parameter Response Meaning
LT none LTL2000 Test Log Dump: 1997 Mar 27 10:01:03
3 entries 98.80% free
Date Time Rl Lamp Idle Load Status Mode
Y:M:D H:M:S mcd/m
2
/lx On Off [v] [v] # [Z,C,T]
1997-03-27 10:00:34 -1 12.0 9.5 12.2 11.3 0 Z
1997-03-27 10:00:45 146 162.9 9.6 12.2 11.3 0 C
1997-03-27 10:01:03 146 162.3 9.6 12.2 11.3 0 T
*
Test log dump date and time
Log statistics
Log header
Data units
Rl zero
Rl calibration
Rl test measurement
LWT
LWF
none Log Full Warning Display On
Log Full Warning Display Off
Enable the Log Full warning in LCD
Disable the Log Full warning in LCD
LX none Rl Test log empty Rl test log clear message
MC none
new count
Averring Measurement count 1000
New value for averring
Number of single measurements used to build an average
signal for calculation of Rl
New value
The instrument is now un calibrated !
OT none
59
180
Automatic off timer = 120 sec.
Automatic off function disable
Automatic off timer = 180 sec.
Automatic turn off time when not in use. Range 60 to 600
sec. Disable for time < 60 sec
RC none
Y
Calibrate Instrument? [Y/N]
Calibrating Instrument!
**** Calibration ****
Lamp On: 163
Lamp Off: 10
Stray Light: 0
On-(Off+SyLeak): 151
Now use RN n command to set new normal value
Confirm calibration with <Y>
Perform calibration measurement
Signal with lamp on
Signal with lamp off
Stray light signal
Compensated signal
Prompt the user to execute the RN command to enter the
correct Rn Value