DELTA LTL2000 Retrometer 25
LTL2000 command set
Command Parameter Response Meaning
RL none *** Rl Measurement ***
1997 Mar 27 10:00:59
Rl: 146 (mcd/m
2
)/lx
Status Code: 0 00000000
Performs a normal Rl measurement. Displays measurement
date and time, the Rl result and
the Status code
RN none
150
(Rl Normal)
Rl_Factor = 2.312
Rl_Normal = 138
Rl_Factor = 2.151
Rl_Normal = 120
Calculated Rl factor
Rl calibration value
Calculated Rl factor based on the new Rl normal value
entered 150 mcd/m
2
/lx
RT none *** Test Measurement ***
1997-Mar-27 10:01:03
Rl =146 (mcd/m
2
)/lx
Signal: 7.3%
Stray Light: 0.0%
VBat Lamp Off: 12.18 V
VBat Lamp On: 11.28 V
Status Code: 0 00000000
Test header date and time
Date and Time
Measured Rl result
Measured signal
Measured stray light signal
Battery voltage with min. load
Battery voltage with max. load
Actual Status code
RZ none *
*** Zero Reading ***
Lamp On: 12
Lamp Off: 10
New Rl_Sy_Leak: 2
Perform zero measurement
Signal with lamp on
Signal with lamp off
New SyLeak signal
SD none Actual Status Code: 0 Bit Pattern: 00000000
Status breakdown:
Actual Status Code: 16 00010000
Status code and bit pattern
if status not zero then the status code is decoded
This example shows a status value of 16. It decodes to