DELTA LTL-X Retrometer 11
Mount the reference calibration unit and perform a complete traceable calibration. Replace the
reference calibration unit with the control calibration unit and make a normal R
L
measure-
ment. Label the Control Calibration unit with the value and the date.
Warnings and errors
An error or warning icon will appear in the upper row of the display if the LTL-X detects a
problem. Press the UP button ▲ to get a description of the most serious error or warning.
Now press the OK button to display a total list of all errors or warnings related to the meas-
urement.
Sound setting
Press the MENU button and select SETTINGS / SOUND. Select KEY CLICK or SOUNDS to
set the individual sound levels for key click and warning/error sounds. Use the UP ▲ or
DOWN ▼ button to set the level. Accept the setting by pressing the OK button. A loud ad-
ditional beep can be selected by pressing the OK button when BEEP is highlighted.
Miscellaneous
The HOME screen can be activated at any time by pressing the HOME button .
Reset log: Press the MENU button and select LOG / CLEAR DATA. Now
select from the menu: LAST, ALL or SERIES.
Date and time: Press the MENU button and select SETTINGS / DATE & TIME. Use the
UP ▲ and DOWN ▼ buttons to set the time and date. Accept the
setting by pressing the OK button.
Power save: Press the MENU button and select SETTINGS / AUTO OFF. Use
▲and ▼ to edit the auto turn off time.
Data exchange / communication
The RSC program, developed by DELTA for use on a PC, allows data to be exchanged be-
tween the LTL-X and a PC. See RSC-program, pg. 40.