16 LTL-X Retrometer DELTA
Notes on error sources
Stray light can occasionally enter the instrument but will be insignificant under normal meas-
urement conditions. Before each measurement, the LTL-X automatically evaluates the leak-
age and compensates for it before the readout. In case of a significant leakage level, a warning
or error message is given and special precautions may be necessary.
Instrument leak, drift and offset errors are compensated by means of data obtained during the
calibration procedure. It is very important to keep the light trap, the dust-protection window
and the ceramic on the calibration unit clean.
The LTL-X illumination angle is 1.24° relative to the road surface. Because of this small an-
gle accurate placement on the road is important. Avoid pebbles and abnormal irregularities.
The LTL-X must be parallel and in contact with the marking surface.
The LTL-X retrometer is a rugged instrument, but it is an optical instrument and must be han-
dled as such.
The LTL-X is factory calibrated. Nevertheless start measurements with a calibration. Study
the display for any warning or error icons. See also Section 4 - Maintenance
Note
High temperature conditions.
Display
If the display is exposed to intense direct sunlight during a longer period of time the display
could become overheated.
To reduce heat an IR-reflecting filter is mounted on the display. It is recommended to close
the protective display shield. The shield also protects against damages and scratches.
"Daylight readable" displays are vulnerable to high temperatures. High temperature will de-
crease the display service life.
When the display temperature exceeds 50C / 122F an audible alarm will sound. The alarm
will sound for 20 second and then pauses for one minute. This sequence continues until the
temperature is lowered or the instrument is turned off. The alarm can be suspended for one
minute by pressing any key.
Battery
The battery is rated to maximum 45C / 113F operating temperature.
Keep the light trap, dust-protection window and ceramics
on the calibration unit clean.
Keep the battery fully charged. A well charged battery is
more resistant to aging and damage.