44 LTL-X Retrometer DELTA
The reference
Zero Signal
Calibration unit
Reference
The road marking is simulated by a piece of white ceramic (the reference) mounted on an
aluminum profile. Ceramics have very stable optical properties because of the smooth surface.
Figure: 5 - Calibration normal
To make sure that calibration of the retrometer is correct it is important that the ceramics and
light trap on the calibration unit is clean and undamaged. Always keep the calibration unit
well protected.
If the ceramic is stained, scratched or broken, the calibration unit has to be replaced and cali-
brated. In case of dust on the ceramics surface of the traceable reference, the use of com-
pressed air is recommended for removal. To clean the ceramic reference on the field calibra-
tion unit, the use of a soft damp cloth is recommended if compressed air fails to remove the
dirt. If necessary, use a mild household detergent. A calibration transfer must always be com-
pleted after the field reference is cleaned. It is necessary to have the traceable reference avail-
able to perform the transfer prior to cleaning the field reference.
To ensure reliable measurements, it is recommended that the calibration unit be periodically
recalibrated to a traceable standard. DELTA Light & Optics offers calibration traceable to
PTB (Physikalsich-Technishe Bundesanstalt). For information contact your distributor or
DELTA.