46 LTL-X Retrometer DELTA
Reference calibration
Now the instrument is ready for the next step and the display will tell you to rotate
the calibration unit so that the white face is facing the instrument tower. Again it is
important that the pins on the side of the unit fit into the slots in the LTL-X. Press OK
to begin the reference calibration. Check the value displayed and if necessary adjust
the calibration value shown in the display, so it matches the value stamped on the
calibration unit.
Press the OK button to perform the calibration.
The calibration procedure is now complete. Remove the calibration unit and store it
properly. Press OK to return to the measurement display.
The instrument automatically compensates for zero signal, leakage and other known errors,
and calculates a calibration factor. This process is fully automatic. If the calibration routine is
followed precisely the retrometer will now display 'true' Rl.
Always store the reference calibration unit in a dry and clean environment.
Figure: 7 Placing the calibration normal.