Delta CS24D User Manual Page 32

  • Download
  • Add to my manuals
  • Print
  • Page
    / 34
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 31
32
Time: There is a certain amount of reading variation over time, typically a few tenths of
a percent per week or less. This effect is compensated for in the CS24D.
12. Troubleshooting
The CS24D's indicator LEDs provide helpful troubleshooting information. The output
LEDs use binary coding to indicate which channel is active. (In Teach mode, the Output
LEDs show the output being taught, not the actual output state). The chart below shows
the operation of the indicator LEDs on the CS24D:
Condition
RUN LED
Output LEDs
No Power
Off
Off
Run mode, No Match
Steady Blue
ALL Off
Run mode, Match
Steady Blue
Steady Green shows Match Output
Teach mode, No Output Selected
Blinking Blue
ALL Off
Teach mode, Output Selected, No
Teach commands
Blinking Blue
Blinking Green shows Selected Sort Output
Teach mode, 1 - 7 Teach commands
Off
Blinking Green shows Selected Sort Output
Teach mode, 8 Teach commands
Blinking Blue
Steady Green shows Taught Sort Output
(The blink rate of the LEDs is about 2 times per second).
If the CS24D is not operating correctly, check for:
1. Dirt buildup on the window
2. Incorrect sensor to target distance
3. Problems with power supply or wiring
Use ColorSense to view data and see how close the match is. If the CS24D seems to be
operating correctly and is still misreading target colors, the sensor may need to be
retrained. (See Teaching section).
13. Support
Delta offers extensive telephone support on all of its products, both before and after
purchase. To have samples tested, send them to:
On new applications, this testing is typically done at no charge. If required, training and
field support can be provided on a time and expense basis.
14. Repairs
If a CS24D needs repair, call Delta to receive an RMA number before returning unit.
Including a brief description of the problem will help to speed the repair time.
Delta Computer Systems, Inc.
1818 SE 17
th
Street
Battle Ground, WA 98604
Attn: Color Sample Testing
(360) 254-8688
FAX (360) 254-5435
Page view 31
1 2 ... 27 28 29 30 31 32 33 34

Comments to this Manuals

No comments