DELTA LTL-XL Retroreflectometer 41
Calibration
The LTL-XL is factory calibrated and very stable but a calibration should always be carried
out before starting a new series of measurements. The instrument is supplied with a calibra-
tion unit to carry out a fast and easy calibration.
Back Front
Calibration procedure
Before mounting the calibration unit check the R
L
and Qd values stamped on the cal-
ibration unit. Place the instrument upon the calibration unit and pull back until the
pins at the back and front of the LTL-XL are in place, see figure 7. Please be very
careful when you place the instrument on the calibration plate not to damage the
calibration unit and / or the light diodes
Press the CALIBRATION button once. Select CALIBRATE R
L
or CALIBRATE QD.
Continue the calibration process applying the following steps
Check the value displayed and, if necessary, adjust the calibration value shown in the
display so it matches the value stamped on the calibration unit.
Figure 7: Placing the calibration unit