MS-250 Mixed Signal Oscilloscope Option
Minimum Voltage Swing (Hysteresis)
You can define the minimum high voltage and maximum low voltage levels using the
Hysteresis controls.
l Minimum high voltage may be set up to 1.4 V above the threshold.
l Maximum low voltage may be set up to 1.4 V below the threshold.
The minimum signal swing is 100 mV. The indeterminate range of 50 mV around the
threshold voltage level is the level below which the MS-500 will not operate. However, the
MS-250 can support a signal as low as 100 mV only if the input signal’s quality is
adequate.