User Interface DF-745 53
7.4.7.5 Temperature Range Alarm Setup
Figure 51: Temperature Alarm Setup
The system is constantly monitoring the ambient temperature in the analyzer cabinet. If
enabled on the Temperature Alarm Setup screen, an alarm can be assigned to trip if the
ambient temperature exceeds preset limits. The user may assign the temperature alarm to
one of four relays.
Use the Next key to move from between fields and use the arrow keys ( and to
change the highlighted selections and to enter numerical values. When done, use the Next
key to move to the Accept button and hit the Enter key to return to the main display.
Using the ESC key at anytime will exit the screen making no changes and return to the
main display.
7.4.7.6 Pressure Alarm Setup
Figure 52: Pressure Alarm Setup
The system is constantly monitoring the pressure in the gas path and the result is displayed
on the front panel. If enabled on the Pressure Alarm Setup screen, an alarm can be assigned
Page view 56